Section 20-Battery Input Board; Replacing The Battery Input Board - Invacare Solo2 TPO100B Internal Repair Manual

Portable oxygen concentrator
Hide thumbs Also See for Solo2 TPO100B:
Table of Contents

Advertisement

20 Battery Input Board

20.1 Replacing the Battery Input Board

1.
Disassemble the Solo2. Refer to Accessing Internal Components on
page 40.
2.
Remove the P.C. Board.
3.
Remove the Sieve Bed Unit.
4.
Remove the compressor box assembly and DO NOT remove any
parts from inside it.
5.
Using a T-10 remove the screws that secure the battery input board
to the rear shroud.
6.
Disconnect the wire harnesses on the existing battery input board
and install on the replacement battery input board.
7.
Attach the battery input board with harnesses to the rear shroud.
8.
Tighten the screws.
Part No 1176122
63
Invacare® Solo2™ Portable Oxygen Concentrator
20 BATTERY INPUT BOARD
Rear Shroud
Battery Input Board
Battery Input Board
Battery
Communication Cable
Power to P.C. Board
Supply Cable