14.2 Installing the P.C. Board

CAUTION
Before handling the P.C. Board, ensure you are properly grounded to prevent static discharge and damage to the components of the P.C.
Board. A static strap MUST be worn and properly grounded using an alligator clip. Electrical conduit or a water pipe can bue used if a
ground is unavailable. Care should be taken to ensure that the alligator clip contacts bare metal surface.
Required tools: Torx® T15 screwdriver
For this procedure refer to Figures 5, 6, and 7.
1.
Place board into the top of the motor sound box ensuring that the tabs line up and also the screw hole.
2.
Install the three Torx® screws that secure the board to the sound box as shown in Figure 5.
3.
Attach the wire connectors to the P.C. Board as shown in Figure 5.
Display Ribbon Cable
Screw Holes
Battery
communication
Valve wire
connectors
External Power
Connector
Part No 1176122
Sound Box Fans
P.C. Board Fan
FIGURE 5
51
Compressor Connector
Battery Connector
Screw Hole