SECTION 5—TROUBLESHOOTING AND MAINTENANCE

Error Codes

ERROR CODE
Common Alarms
256
512
768
1024
1280
1536
1792
2048
2049
2304
2560
3584
3840
4352
4608
4096
4097
4098
4099
.
ERROR CODE
Compressor Alarms
3328
3329
3330
Alarm - Power not on but compressor told
3331
SOLO
2
DESCRIPTION
Warning - Battery less than 25%
Warning - Battery less than 15%
Warning - No Breath Found
Warning - Battery less than 10%
Warning - Breath Rate High
Alarm - Unit Temp HI/LO
Warning - Start Temp HI/LO
Alarm - Battery Temp HI/LO
Alarm - Battery Temp HI/LO
Warning - Low O2 Purity
Alarm - Low O2 Purity
Alarm - Stuck Button
Alarm - Low Flow Output
Warning - EXT PWR Low
Alarm - EXT PWR Low
Alarm - Fan is not moving
Alarm - High pressure
Alarm - Low pressure
Alarm - Max shift time exceed
DESCRIPTION
Alarm - Compressor lock
Alarm - Starting speed check
Alarm - Power on but compressor
shutdown
to start
20
COMMENT
Charge battery, see Owner's Manual
Charge battery, see Owner's Manual
Check Canula
Charge battery, see Owner's Manual
Reduce Activity
Check ambient temp and blockage of air
inlet/outlet
Check ambient temp and blockage of air
inlet/outlet
Check ambient temp and blockage of air
inlet/outlet
Check ambient temp and blockage of air
inlet/outlet
Check ambient temp and blockage of air
inlet/outlet
Check ambient temp and blockage of air
inlet/outlet
Remove power, restart unit
Check Canula and patient outlet filter
See Owner's Manual
Find New Power
Operating Retry - Call Provider
Operating Retry - Call Provider
Operating Retry - Call Provider
Operating Retry - Call Provider
COMMENTS
3328
3329
3330
3331
Part No 1164898