Invacare Solo2 TPO100B Internal Repair Manual page 45

Portable oxygen concentrator
Hide thumbs Also See for Solo2 TPO100B:
Table of Contents

Advertisement

Replacing
When replacing batteries because of problems with unit performance, replace all three batteries. DO NOT mix used and new
batteries.
1.
Place battery "1" into the shroud and connect to Connector "1".
2.
Place battery 2 into the shroud and connect to Connector "2".
3.
Place battery 3 into the shroud and Connect to Connector "3".
4.
Take all wires and carefully push them down into the slot on the side of the battery box.
5.
Ensure that all wire after pushing down into the slot are not above the surface and place the cover. On the other half of the shroud and screw
the cover back together using the T-10 Torx® screwdriver.
6.
Replace the Battery label and serial number.
New batteries MUST be fully changed before use.
DETAIL "A"
3
Part No 1176122
2
1
Wires in place in battery box
DETAIL "B"
3
45
Invacare® Solo2™ Portable Oxygen Concentrator
13 BATTERY PACK