Replacing Batteries In Battery Pack - Invacare Solo2 TPO100B Internal Repair Manual

Portable oxygen concentrator
Hide thumbs Also See for Solo2 TPO100B:
Table of Contents

Advertisement

13 BATTERY PACK

13.3 Replacing Batteries in Battery Pack

Removing
CAUTION
When disassembling the battery pack make sure that no exposed connectors touch each other, otherwise, shorts of the individual batterys
may occur.
Required tools: Torx® T10 screwdriver
If the battery pack does not charge or the Solo2 does not power up in the required time as stated in Product Parameters, replace
all battery packs.
1.
Remove the label from the back of the battery pack (Detail "A") to expose the screws (Detail "B").
2.
Remove and retain the screws. Remove the back cover.
If the battery label rips during removal, remove the serial number label and replace the battery label with a new one. Apply the serial
number label to the replacement label.
3.
Unplug each battery from the bottom connector (Detail "C").
DETAIL "A"
Invacare® Solo2™ Portable Oxygen Concentrator
DETAIL "B"
Serial Number
Label
Battery Pack
Label
44
DETAIL "C"
Screws
Battery Pack with
Label removed
Batteries
Battery Pack
Connector
Part No 1176122