Sharp MX-FXX1 Service Manual page 22

Digital full color multifunctional system option facsimile expansion kit
Hide thumbs Also See for MX-FXX1:
Table of Contents

Advertisement

66-15
Purpose
: Adjustment
Function (Purpose) : Used to execute the dial pulse (20PPS)
send test and to adjust the make time.
Section
: FAX
Item
: Operation
Operation/Procedure
1)
When the machine enters Simulation 66-15, the following
screen is displayed.
* Enter the make time (0 to 15) with 10-key and press [OK]
button, and the entered make time is temporality saved.
* Enter the make time (0 to 15) with 10-key and press [EXE-
CUTE] button instead of [OK] button, and [EXECUTE] but-
ton is highlighted and the dial pulse is sent from the line in
the set make time.
2)
When [EXECUTE] button is pressed, it is highlighted and the
dial pulse is sent from the line in the set make time.
* The dial pulse in this example is up to 20 digits registered
with SIM66-13.
3)
To end the dial test, press [EXCUTE] button again. The button
returns to the normal display and the test is terminated.
Also when dialing the number is completed, [EXECUTE] but-
ton returns to the normal display and the test is terminated.
SIMULATION NO.66-15
TEST
DIAL TEST(20PPS)
A
4
MAKE TIME [+9ms]
A:
4
0
15
66-16
Purpose
: Adjustment
Function (Purpose) : Used to execute the DTFM signal send test
and to adjust the send level.
Section
: FAX
Item
: Operation
Operation/Procedure
1)
When the machine enters Simulation 66-16, the following
screen is displayed.
* Item A (High group frequency) and item B (High-Low group
frequency) are selected with [↑] [↓] buttons. When the
selected item is the top (last) item, [↑] [↓] buttons gray out.
(The highlighted section of the set value is switched and dis-
played on the set setting area.)
When [↑] [↓] buttons gray out, they are disabled.
* Enter the High group or High - Low group (0 to 15) with 10-
key and press [OK] button. The set sign send level is tempo-
rality saved.
* Enter the single send level (0 to 15) with 10-key and press
[EXECUTE] button instead of [OK] button, and [EXECUTE]
button is highlighted and the dial pulse is sent from the line
by the setting of high group or low group.
2)
When [EXECUTE] button is pressed, it is highlighted and the
dial pulse signal is sent from the line by the setting of high/low
group of the signal send level.
3)
To terminate the dial test, press [EXECUTE] button. The button
returns to the normal display and the test is terminated.
When dialing the dial number is completed, [EXECUTE] button
returns to the normal display and the test is terminated.
CLOSE
EXECUTE
OK
MX-FXX1 SIMULATIONS, FAX SOFTWARE SWITCH 5 – 13
SIMULATION NO.66-16
TEST
DIAL TEST(DTMF)
A
4
HIGH(dB)
A:
4
B
4
HIGH-LOW
0
15
66-17
Purpose
: Operation test/Check
Function (Purpose) : Used to send the DTMF signal to the line
and the speaker. (Send level: Max.)
Section
: FAX
Item
: Operation
Operation/Procedure
1)
When the machine enters Simulation 66-17, the following
screen is displayed. (Default, left upper selected)
2)
When a button of a send signal is selected, it is highlighted and
the previously set button returns to the normal display.
* Only one kind of signal to be sent can be selected.
3)
When [EXECUTE] button is pressed, it is highlighted and sig-
nals are sent.
* When another signal send button is pressed during sending
signals, the send signal can be changed.
4)
To stop signal sending:
When [EXECUTE] button is pressed, it returns to the normal
display and signal sending is interrupted.
SIMULATION NO.66-17
TEST
DTMF SIGNAL OUTPUT.(LEVEL MAX)
2
1
4
5
7
8
0
66-18
Purpose
: Operation test/Check
Function (Purpose) : Used to send the DTMF signal to the line
and the speaker. (Send level: Soft SW set-
ting)
Section
: FAX
Item
: Operation
Operation/Procedure
1)
When the machine enters Simulation 66-18, the following
screen is displayed.
2)
When a button of a send signal is selected, it is highlighted and
the previously set button returns to the normal display.
3)
When [EXECUTE] button is pressed, it is highlighted and sig-
nals are sent.
* When another signal send button is pressed during sending
signals, the send signal can be changed.
4)
To stop signal sending:
When [EXECUTE] button is pressed, it returns to the normal
display and signal sending is interrupted.
CLOSE
EXECUTE
OK
3
6
9
#
EXECUTE