5— Getting Started - Making Initial Checks
Making Initial Checks
Checking configuration data
Before doing anything else, it is a good idea to review the transmitter's factory-set configuration
parameters. Table 16 outlines the steps.
Step
1
From the "Online" menu, enter "Device setup" by pressing the right arrow ( ⇒ ) key on the
communicator keypad.
2
Press the down arrow ( ⇓ ) key to scroll down to menu-item "5 Review". When highlighted press
the right arrow ( ⇒ ) key to enter review function. A display similar to the one shown below
appears.
3
Press PREV and/or NEXT to scroll through and view the configuration data, including:
•
Manufacturer*
•
Transmitter model*
•
Transmitter Measurement type*
PV unit
•
•
Maximum and minimum range limits
- PV URL
- PV LRL*
•
PV Damping
•
PV output in % of range
•
PV transfer function (Output conformity)
PV upper range value (URV) and lower
•
range value (LRV)
PV AO (analog output) in milliamperes
•
•
PV AO alarm option (failsafe direction)**
•
SV (Secondary Variable) unit
* This information is fixed and cannot be changed by reconfiguring the transmitter.
** Alarm option and transmitter security are jumper-selectable on the electronics board.
4
After reviewing the transmitter data, press EXIT which takes you back to the "Device setup"
display.
42
Table 16 Reviewing Factory-Set Configuration Parameters
ST3000:PT 3011
Review
Manufacturer
Honeywell
HELP
ST 3000 HART Transmitter Release 300 User Manual
Action
PREV
NEXT
EXIT
•
PROM ID*
•
Tag name
•
Date
Descriptor
•
•
Message
•
Write protection**
Final assembly number
•
•
Device ID*
•
Universal revision*
Field device revision*
•
•
Software revision*
•
Polling address
Number of request preambles*
•
10/05