Diagnostic Messages; Table 48 Summary Of Diagnostic Messages For Critical Failures; St 3000 Hart Transmitter Release 300 User Manual - Honeywell ST 3000 User Manual

Smart transmitter
Hide thumbs Also See for ST 3000:
Table of Contents

Advertisement

Diagnostic Messages

Summary
The diagnostic messages can be grouped into one of these three categories:
Critical Failures
Non-Critical Failures
Communication Errors
A description of the messages in each category is given in the following paragraphs.
Critical failures
Table 48 summarizes the critical communicator status message displays. A critical failure drives the
transmitter's output to its failsafe direction—upscale or downscale.
INVALID DATABASE
CHAR PROM FAULT
SUSPECT INPUT
DAC DIODE FAULT
NVM FAULT
RAM FAULT
PROM FAULT
PAC FAULT
Once a critical fault has been corrected, you must clear the critical status from the transmitter. See Clearing
Critical Status later in this section.
10/05

Table 48 Summary of Diagnostic Messages for Critical Failures

Message

ST 3000 HART Transmitter Release 300 User Manual

11— Troubleshooting - Diagnostic Messages
Description
Database corrupted upon power-up.
Characterization PROM failure.
Input pressure may be incorrect
Digital to Analog Converter (DAC) fault
Transmitter Nonvolatile Memory (NVM) fault
Transmitter Random Access Memory (RAM) fault
Transmitter Programmable Read Only Memory
(PROM) fault
Currently not implemented.
114

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents