Non-critical failures
Table 49 summarizes the non-critical communicator status message displays. All communicator functions
remain operational during a non-critical failure.

Table 49 Summary of Diagnostic Messages for Non-Critical Failures

SENSOR OVER TEMP
EXCESS ZERO CORR
EXCESS SPAN CORR
IN OUTPUT MODE
M.B. OVERLOAD OR
METERBODY FAULT
CORRECTS RESET
NO DAC TEMP COMP
Other error messages that may appear due to noncompatability of communicator software or transmitter
communication mode.
NOTICE: Upgrade 275 software to access new
Xmtr functions. Continue with old description?
In multidrop mode
Communication errors
Table 50 summarizes the message displays associated with communication errors. All communicator
functions are disabled when a communication error occurs.

Table 50 Summary of Diagnostic Messages for Communication Errors

Device Disconnected
No Device Found
10/05
Message
Message
Message
ST 3000 HART Transmitter Release 300 User Manual
11— Troubleshooting - Diagnostic Messages
Description
Meter body temperature is too high.
Zero calibration value is too large (shift is larger
than characterization).
SPAN correction factor is outside the acceptable
limits for accurate operation.
Transmitter is operating as current source.
Input pressure is more than two times greater
than the Upper Range Limit of the transmitter.
Must recalibrate transmitter to attain required
accuracy.
No temperature compensation data exists for
calculations
Description
You have connected to a device that has a newer
revision of device description than what is in the
communicator.
The transmitter poll address is not 0 (zero). You
have tried to change the analog output of a
transmitter that is in multidrop mode.
Description
Communication with a device has been
interrupted.
Communicator was unable to establish
communications with any device upon power-up.