Pressure Or Liquid Level Measurement With Gp Transmitter; Figure 43 Typical Piping Arrangement For Pressure Measurement With Gp Type Transmitter; Figure 44 Typical Piping Arrangement For Liquid Level Measurement With Gp Typetransmitter - Honeywell ST 3000 User Manual

Smart transmitter release 300 and smart field communicator model sts103
Hide thumbs Also See for ST 3000:
Table of Contents

Advertisement

7.8
Pressure or Liquid Level Measurement with GP
Transmitter
Procedure
Figure 43
Typical Piping Arrangement for Pressure Measurement with GP Type Transmitter
Process
Figure 44
Typical Piping Arrangement for Liquid Level Measurement with GP Type
Transmitter
6/05
ST 3000 Release 300 and SFC Model STS103 User's Manual
The procedure in Table 43 outlines the steps for starting up a gauge
pressure (GP) type transmitter in a pressure or liquid level measurement
application. Refer to Figures 43 and 44 for the piping arrangement
identification and Figure 38 for typical SFC and meter connections.
Block-off
valve no.1
Block-off
valve no.2
Block-off
valve
Tap location at the minimum
level to be measured
Pipe
Union
Plug
Tee connector
To Process Head
connection on
meter body
Gauge
Pressure
Transmitter
Gauge
Pressure
Transmitter
Continued on next page