Honeywell ST 3000 User Manual page 123

Smart transmitter release 300 and smart field communicator model sts103
Hide thumbs Also See for ST 3000:
Table of Contents

Advertisement

6.12
Configuring Smart Meter Using Pushbuttons,
Selecting Engineering
Units, continued
Table 33
Selecting Engineering Units, continued
Step
5
If selected engineering unit does not match
one of six unit indicators on meter, peel off
matching stick-on unit label from sheet
(drawing number 30756918-001) and paste it
in lower right hand corner of meter.
6
If you selected Custom or Flow engineering
units, go to Tables 35 and 36 to set lower and
upper display limits for smart meter display.
6/05
ST 3000 Release 300 and SFC Model STS103 User's Manual
Action
Meter Display
Use stick-on label for engineering units
without indicators on display.
V AR
SEL.
%
0
100
1
0
2
.
ANALOG
Kg/cm
Lower and upper display limits have not been
set for Custom or Flow engineering units.
VA R
SEL.
%
0
-
U
L
ANALOG
Continued
UPPE R
VALUE
UNITS
S E T
LOWER
VALUE
2
Stick-on label
identifies selected
engineering units
UPPE R
VALUE
100
UNITS
S E T
FLOW
LOWER
VALUE
GPM
109