Figure 13 Typical Remote Diaphragm Seal Transmitter Installation - Honeywell ST 3000 User Manual

Smart transmitter release 300 and smart field communicator model sts103
Hide thumbs Also See for ST 3000:
Table of Contents

Advertisement

4.2
Mounting ST 3000 Transmitter,
Remote seal
mounting, continued
6/05
ST 3000 Release 300 and SFC Model STS103 User's Manual
Table 11
Mounting Remote Diaphragm Seal Transmitter, continued
Step
3
If Transmitter Model
Number is...
STR93D or STR12D
STR13D
ATTENTION
accommodate the flange extension.
4
It is the End User's responsibility to provide a flange gasket and
mounting hardware that are suitable for the transmitter's service
condition
Figure 13
Typical Remote Diaphragm Seal Transmitter Installation.
LP Side
- Model STR93D
- Model STR12D
HP Side
- Model STR13D
H2
Fixed
Ref. Leg
HP Side
- Model STR93D
- Model STR12D
LP Side
- Model STR13D
Continued
Action
Then Connect Remote
Seal on...
low pressure (LP) side of
transmitter to upper flange
mounting on tank wall for
fixed or constant head H2.
high pressure (HP) side of
transmitter to upper flange
mounting on tank wall for
fixed or constant head H2.
On insulated tanks, remove enough insulation to
Maximum Level
Minimum Level
Variable
Head H1
37

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents