Fluke Versiv Series User Manual page 266

Cabling certification product family
Hide thumbs Also See for Versiv Series:
Table of Contents

Advertisement

Find Quality Products Online at:
Table 9. Settings for HDR OTDR Tests(cont.)
Test Type
DataCenter OTDR: This test is optimized for fiber
installations that have short links, many connections, and
(continued)
the possibility of large reflections.
By default, the DataCenter OTDR test uses
1310 nm for single-mode fiber. This wavelength is
typically used in data centers. You can select
other wavelengths if necessary.
You must use launch compensation when you do
a DataCenter OTDR test.
FaultMap: The FaultMap test can show connections that
do not show on the OTDR EventMap and connections that
are poor because they have high reflectance. See "The
FaultMap Test" on page 279.
SmartLoop OTDR (Auto) and SmartLoop OTDR (Manual):
The SmartLoop test lets you connect the far ends of the
two fibers in a link so that one OTDR test gives you results
for both fibers. The Auto and Manual settings operate the
same as for the OTDR test. See "The SmartLoop Test" on
page 286.
FiberInspector: The FiberInspector test lets you use an
optional FI-7000 FiberInspector
the endfaces in fiber optic connectors.
For all Test Types except FaultMap and
FiberInspector, you can use the SETUP button on
the TRACE screen to change some OTDR settings
manually after a test. See "How to Quickly
Change OTDR Settings After a Test" on page 277.
Manual OTDR
This item shows only if you select Manual OTDR or
Settings
SmartLoop OTDR (Manual) for the Test Type. Manual
OTDR mode lets you select settings to control the test
parameters for the trace. See the Technical Reference
Handbook.
GlobalTestSupply
www.
Use the HDR OTDR
Settings for HDR OTDR Tests
Notes
video probe to inspect
Note
(continued)
.com
sales@GlobalTestSupply.com
257