KION / KION-i
A

@
.

E382 E392E 061 01 02 02
• To remove the pressure regulator REG5–REG7 (8),
unscrew the two screws (9) using a 5 mm
hexagonal wrench.
• When mounting the pressure regulator, always use
new O-rings (10). Apply a thin layer of special
grease on the O-rings. Tightening torque for the
screws (9) is approx. 18 Nm.
• To remove the pressure transducer (11), unscrew
the transducer using a 19 mm open-end spanner.
• When mounting the pressure transducer, always
use a new gasket (12). Apply a thin layer of special
grease on the gasket.
• Tighten the pressure transducer by hand.
Tighten further approx. half a turn using the open-
end spanner.
Siemens-Elema AB
Disassembling and assembling
4 - 25