Disassembling and assembling
1
2
3
4
4 - 22
Front panel controls
• Remove the control panel housing as previously
described.
• Remove the knob cover (1).
• Remove the pin (2) if the control has a push-button
release function.
• Loosen the nut (3) holding the knob and lift off the
knob from the control shaft. Use the knob holder
tool (Order No. 62 04 197 E380E) when removing
and mounting the control knobs.
• Disconnect the connector (4).
4
• Remove the two screws (5) and lift off the control.
The assembling procedure is the reverse of the
disassembling procedure previously performed.
The following points must be noted during
assembling:
5
• If the control knob has a pointer, make sure that:
– On the control panel switches; this knob pointer
aligns with the indication lines on the panel in the
different switch settings.
– On the control panel potentiometers; this knob
1
pointer aligns with the end position indications on
the panel and, if applicable, with the values
showed on the corresponding display.
• Tighten the nut (3) with a torque of approx.:
2
– 150 Ncm on the big knobs
– 90 Ncm on the small knobs.
• If the knob is hard to turn, it can be necessary to
put a 0.2 mm thickness gauge (1) between the
control panel and the knob while mounting the
knob (2).
Siemens-Elema AB