PAN: DIAG_STARTUP_CPU_TEST_FAILED
PAN: DIAG_STARTUP_ROM_TEST_FAILED
PAN: DIAG_STARTUP_RAM_TEST_FAILED
PAN:
DIAG_STARTUP_WATCHDOG_TEST_FAILED
PAN: DIAG_STARTUP_TIMER_TEST_FAILED
PAN: DIAG_STARTUP_RESET_OCCURED
PAN: DIAG_STARTUP_LED_TEST_FAILED
Display-error
PAN: CAN_ERROR_BUFFER_FULL
PAN: DIAG_CAN_DATA_OVERRUN
Fault: Hardware failure.
Description: CPU fails during startup hardware test.
What to do: Replace PC 1672 Control Panel
Fault: Hardware failure.
Description: ROM fails during startup hardware test.
What to do: Replace PC 1672 Control Panel.
Fault: Hardware failure.
Description: RAM fails during startup hardware test.
What to do: Replace PC 1672 Control Panel.
Fault: Hardware failure.
Description: Watchdog fails during startup hardware test.
What to do: Replace PC 1672 Control Panel.
Fault: Hardware failure.
Description: Timer fails during startup hardware test.
What to do: Replace PC 1672 Control Panel.
Fault: Hardware failure.
Description: Reset occurred during the startup hardware test.
What to do:
1.
Check +5V power supply to PC 1672 Control Panel
2.
Replace PC 1672 Control Panel.
Fault: Hardware failure (display-error).
Description: One of the Control panel LEDs fails during the startup hardware test.
What to do: Replace PC 1672 Control Panel.
Fault: Communication bus (CAN) is not working properly.
Description: The circular CAN receive buffer is full, at least one CAN-message is
missed. Circular buffer holds 10 CAN-messages.
What to do:
1.
If this log is repeted several times; check CAN communicating units, i.e.
microprocessor controlled PC boards, multigas analyzer, KION Monitor and
DuoView. Also check PC board- and cable connectors.
2.
Report to EM-HSC.
Fault: Communication bus (CAN) is not working properly.
Description: Data overrun has occurred, at least one CAN-message is missed.
What to do:
1.
If this log is repeted several times; check CAN communicating units, i.e.
microprocessor controlled PC boards, multigas analyzer, KION Monitor and
DuoView. Also check PC board- and cable connectors.
2.
Report to EM-HSC.