ALA:
DIAG_LOW_BATTERY_STARTUP_VOLTAGE
ALA: DIAG_ERROR_LINE_ERROR
ALA:
DIAG_MEASURE_NOT_RESP_RESEND_ALA
RMS
ALA:
DIAG_CONTROL_NOT_RESP_RESEND_ALA
RMS
ALA:
DIAG_PANEL_NOT_RESP_RESEND_ALARM
S
Fault: Failure in the internal battery charging voltage.
Description: The charging voltage is too low (when running on mains power) during
startup hardware test.
What to do:
Check if +24V_PTC is available at P59 on PC 1675 Power Supply:
• If +24V_PTC is not available, replace the AC/DC Converter.
Note: As the two fans inside the battery drawer runs on +24 V, this error is easy to
identify by checking the function of these fans.
• If +24V_PTC is available, this voltage is correctly generated by the Power Supply
system. The error must be found in the cables/connectors or on PC 1672 Control
Panel.
Fault: Hardware failure during the startup test of sound generation and error line.
Description: The Emergency line activates/deactivates the buzzer by checking that
Sound Sense is activated/deactivated. If not, this diagnostic and
DIAG_ALARM_BUZZER_ERROR (032) is written. Technical alarm and alarm 830 is
generated.
What to do:
1.
Replace PC 1672 Control Panel
2.
Replace PC 1737 Measuring. Check if any other adjacent diagnostic log can
indicate this possible error.
3.
Replace PC 1730 Control. Check if any other adjacent diagnostic log can
indicate this possible error.
Fault: MEA is not responding.
Description: No acknowledge is received from this node on the RESEND_ALARMS
message within 1 second. (Warm start). Technical alarm is generated.
What to do:
1.
Check that the MEA-PROM is correctly mounted
2.
Replace PC 1737 Measuring.
Fault: CON is not responding.
Description: No acknowledge is received from this node on the RESEND_ALARMS
message within 1 second. (Warm start). Technical alarm is generated.
What to do: Replace PC 1730 Control.
Fault: PAN is not responding.
Description: No acknowledge is received from this node on the RESEND_ALARMS
message within 1 second. (Warm start). Technical alarm is generated.
What to do:
1.
Check that the PAN-PROM is correctly mounted
2.
Replace PC 1672 Control Panel.