Siemens KION Service Manual page 170

Anesthesia workstation
Table of Contents

Advertisement

Maintenance
7
7 - 4
Performing the Twelve-months
maintenance
• Disassembling and assembling of the workstation
is required when replacing parts included in the
"Maintenance kit 12 months for KION". If not
stated otherwise, refer to chapter "Disassembling
and assembling" for instructions.
• The letters A – V in the text below refers to the
description of the Maintenance kit on page 7 - 3.
• Prepare the unit as described in section
"Preparations" above in this chapter.
Six-months maintenance
The Six-months maintenance is a part of the Twelve-
months maintenance. Perform as described in the
"KION / KION-i – Operating Manual", chapter "Six-
months maintenance".
Replace:
A. Fresh gas valve tube and PEEP valve tube.
B. Bacteria filter with tube for the PEEP pressure
transducer.
C. Bacteria filters for the silicone muff in the patient
cassette docking station.
D. Expiratory flow transducer screen (mesh net)
including screw.
Clean the following parts before assembling:
• Fresh gas valve tube and PEEP valve tube
• Fresh gas flow transducer and Expiratory flow
transducer.
• Silicone rubber connections for the Fresh gas flow
transducer.
• Exhaust gas hose connected between the PEEP
valve and the EVAC outlet.
For cleaning instructions, refer to the "KION /
KION-i – Operating Manual", chapter "Routine
cleaning".
Gas modules
Replace:
E. Filters in the gas modules. When replacing filter,
move the rubber seal from the old to the new
filter.
E
Siemens-Elema AB
KION / KION-i
E382 E392E 061 01 02 02