Siemens KION Service Manual page 138

Anesthesia workstation
Table of Contents

Advertisement

... 26 AE
CAL_ERR_START_P_TOO_LOW
... 26 AF
CAL_ERR_NOT_ENOUGH_P1_SAMPLES
... 26 B0
CAL_ERR_NOT_ENOUGH_P2_SAMPLES
... 26 B1
CAL_ERR_P_HIGHER_AFTER_DELAY
... 26 B2
CAL_ERR_START_P_TOO_HIGH
... 26 B3
CAL_ERR_START_P_TOO_LOW_2
... 26 B5
CAL_ERR_P_NOT_RAISE_ON_REFILL
x
262
...01 06 ...
CON:
DIAG_PUC_AUTOVENT_NOT_ACCEPTABLE
_LEAKAGE
Sub data:
... AB CD
LEAKAGE_MORE_THAN_150_ML
Fault: PuC technical error code 9902. Large leakage, disconnect.
The volume could not be calculated, the test pressure was not reached in 15 seconds.
The fresh gas filling flow was 20 l/min.
What to do: Refer to section "If the Pre-use check fails".
Fault: PuC technical error code 9903.
The Start pressure could not be calculated as a mean value because the number of
samples was less than 2 in 0.5 seconds.
What to do:
1.
Power cycle the system.
2.
Replace PC 1730 Control
Fault: PuC technical error code 9904.
The End pressure could not be calculated as a mean value because number of
samples was less than 2 in 0,5 seconds.
What to do:
1.
Power cycle the system.
2.
Replace PC 1730 Control
Fault: Pre-use check technical error code 9905. Gas leakage into the workstation.
The end pressure is more than 1 cmH2O above the start pressure. This could be due
to a leakage into the system from a gas module, a mushroom valve, the O2-flush or
emergency gas valves or from a disconnected sampling line. If the end pressure is
above the start pressure, but the difference is less than 1 cmH2O, the system is
assumed to be tight.
What to do: Refer to section "If the Pre-use check fails".
Fault: PuC technical error code 9906. The workstation is not evacuated.
The system is not evacuated in 10 seconds after the last "pressure puff" before the
ramp for volume calculation. The pressure is higher than 1.5 cmH2O.
What to do: Refer to section "If the Pre-use check fails".
Fault: PuC technical error code 9907. Leakage 6-19 l/min.
The initiated test pressure 41 cmH2O could not be reached in 5 attempts.
What to do: Refer to section "If the Pre-use check fails".
Fault: PuC technical error code 9909. Gas leakage in the workstation.
The refilling flow is the "calculated leakage + 200 ml" for 5 seconds. If the pressure is
not increasing, then an assumed leakage between the gas modules and the
inspiratory gas block is detected.
What to do: Refer to section "If the Pre-use check fails".
Fault: "Leakage automatic ventilation" test failed during PuC. No error code stated.
Description: The leakage was greater than 150 ml/min but not generating an error
code.
What to do: Refer to section "If the Pre-use check fails".
The leakage (ml/min) is presented in hexadecimal form.