Siemens KION Service Manual page 132

Anesthesia workstation
Table of Contents

Advertisement

... 00 02
... 01 11
... 02 11
... 01 12
... 02 12
... 01 13
... 02 13
... 01 14
... 02 14
... 01 21
... 02 21
... 01 22
... 02 22
... 01 23
... 02 23
... 01 24
... 02 24
... 01 31
... 02 31
... 01 32
... 02 32
... 01 33
... 02 33
... 01 34
... 02 34
... 03 00
... 03 01
... 03 02
x
230
... 00 E6 ...
CON: DIAG_USER_INTERACTION_TIMEOUT
Pre-use check aborted
Switching Breathing system:
... 01 xx = Ventilation mode switch or Standby
... 02 xx = ON
Standby
Circle system, Manual ventilation
Circle system, Volume control
Circle system, Pressure support
Circle system, Pressure control
Non-Rebreathing, Manual ventilation
Non-Rebreathing, Volume control
Non-Rebreathing, Pressure support
Non-Rebreathing, Pressure control
Aux Fresh Gas Outlet, Manual ventilation
Aux Fresh Gas Outlet, Volume control
Aux Fresh Gas Outlet, Pressure support
Aux Fresh Gas Outlet, Pressure control
Calibration passed
Calibration failed
Calibration aborted
Fault: Calibration sequence not fulfilled.
Description: The user has not respond with "calibration continue" when starting a
calibration or accepted/set barometric pressure during calibration (10 minutes
response timeout)
What to do: Restart Calibration.