KION / KION-i
Optional equipment
22. DuoView
For additional monitoring capabilities, an optional
KION D
V
can be connected to the KION / KION-i
UO
IEW
workstation.
The KION D
V
is a Surgical Display designed to
UO
IEW
be an integrated part of the KION / KION-i
workstation.
There are two versions of the KION D
• KION D
V
with S/N 01000 or lower.
UO
IEW
• KION D
V
with S/N 01001 or higher.
UO
IEW
Both versions have the same functionality.
The knob on the D
V
UO
IEW
screen. On D
V
with S/N 01001 or higher, this
UO
IEW
knob can also be used to switch the D
Off.
The KION D
V
contains the following main parts:
UO
IEW
• The Surgical Display Controller. The SDV-SW is
stored in a Flash-PROM in the Surgical Display
Controller. SW downloading is described in chapter
"Service procedures"
• A 10.4 inch color VGA screen.
The D
V
is connected to the KION / KION-i
UO
IEW
workstation with one connector containing 12 V
power supply as well as RS232 communication.
The D
V
will start when the KION/KION-i power
UO
IEW
switch is turned from Off to Standby or On.
During the D
V
start-up sequence, communi-
UO
IEW
cation with KION Monitor will be established and
start-up information will be displayed on the D
The D
V
will be shut off when the KION / KION-i
UO
IEW
power switch is set to Off.
The D
V
is controlled by the KION Monitor. With
UO
IEW
the D
V
connected (and switched ON), the
UO
IEW
information will be displayed as follows:
• KION Monitor: A fixed layout containing only
ventilation and gas monitoring parameters.
• KION D
V
: Hemodynamic and ECG parameters.
UO
IEW
Any parameter setting on the D
the KION Monitor as described in the User's Guide
for the KION Monitor.
E382 E392E 061 01 02 02
V
:
UO
IEW
sets the brightness on the
V
On and
UO
IEW
V
UO
IEW
V
is made on
UO
IEW
Siemens-Elema AB
23. Multigas analyzer
An optional multigas analyzer can be installed into the
KION / KION-i Workstation. There are two different
multigas analyzers that can be used:
• KION MultiGas 2000
• KION MultiGas+.
A sampling tube is connected to the sampling
adapter at the Y-piece. A gas pump in the multigas
analyzer takes gas samples from the Y-piece. The gas
sample is analyzed by the multigas analyzer and the
measured values are shown on the KION Monitor.
The gas sample is then returned to the expiratory
channel in the P
ATIENT
F
T
. Thus, the measured expiratory flow
LOW
RANSDUCER
includes the gas sample.
If the temperature in the battery drawer compartment
is above 45°C (113°F), e. g. due to B
F
failure, the function of the multigas analyzer may
AN
be affected.
For technical information regarding the multigas
analyzers, refer to the Service Manual for the
concerned unit (KION MultiGas 2000 or KION
MultiGas+).
24. Isolation transformer
The purpose of the optional KION I
T
is to provide the possibility for mains
RANSFORMER
power connection of auxiliary equipment while
maintaining the total system earth leakage current
below the limits in IEC 60601-1. It provides galvanic
isolation from the wall mains outlet.
.
The I
T
SOLATION
RANSFORMER
outlets; 115 V or 230 V, 50 or 60 Hz. Max. outlet is
500 VA and max total leakage current is 100 µA.
A green lamp on the transformer indicates when
mains power is available.
The I
T
SOLATION
RANSFORMER
• Two resetable fuses on the primary side:
– 6 A on the 115 V transformer.
– 3 A on the 230 V transformer.
• Two 5x20 mm UL approved fuses on the
secondary side:
– 5 A / 250 V slow-blow on the 115 V transformer.
– 2.5 A / 250 V slow-blow on the 230 V
transformer.
The mains power outlets are designed according to
the European standard IEC 320/C14.
For further information, refer to the "KION Optional
Accessories Operating Manuals".
Description of functions
C
prior to the E
ASSETTE
XPIRATORY
D
ATTERY
RAWER
SOLATION
contains four mains power
is equipped with four fuses:
3
3 - 25