6. Inspiratory control
Controls the gas delivered to the breathing system.
The main functions are:
• The three G
M
. They are basically of the
AS
ODULES
same design and function as the gas modules
(inspiratory valves) in the Servo Ventilator 300.
• Five safety valves, SV1 – 4 and SV8.
• F
G
P
C
RESH
AS
RESSURE
ONTAINER
pressure for the fresh gas supply.
• Inspiratory fresh gas channel that measures and
regulates the fresh gas to the breathing system.
Gas modules
There are two different versions of the gas module:
• Type I recognized by the metal nozzle unit.
• Type II recognized by the plastic nozzle unit.
As from S/N 01501, Type I was replaced by Type II in
the production of KION workstation. For information
regarding compatibility between Type I and Type II,
refer to SpeedInfo 008/02 EM LSS.
If not stated otherwise, this functional description is
valid for both versions.
The three G
M
are:
AS
ODULES
• G
M
AIR (N
O) – BP supplies AIR or N
AS
ODULE
2
the V
M
.
APORIZER
AGAZINE
• G
M
O
– BP supplies O
AS
ODULE
2
M
.
AGAZINE
• G
M
D
G
AIR (O
AS
ODULE
RIVE
AS
as drive gas to the R
EBREATHING
The G
M
marked BP are modified to handle
AS
ODULES
the backpressure created when supplying fresh gas
to vaporizers. Max. flow from the gas modules
modified for backpressure is reduced to 2 l/s.
The modification concerns PC boards inside the gas
module and the nozzle unit. Without these modifica-
tions, the backpressure would disable the
diaphragms closing function.
The G
M
are pneumatically connected to the
AS
ODULES
I
G
B
and electrically connected to
NSPIRATORY
AS
LOCK
PC 1765 V
I
ALVE
NTERCONNECTION
E382 E392E 061 01 02 02
to create the working
O to
2
to the V
APORIZER
2
) supplies AIR or O
2
2
S
.
ECTION
.
Siemens-Elema AB
The G
M
are factory calibrated.
AS
ODULES
The adjustment potentiometer located at the top
of each G
M
(Type I) must not be
AS
ODULE
adjusted.
Each G
M
must not be disassembled
AS
ODULE
further than described in chapter "Maintenance".
4
5
6
1. Filter
2. Inspiratory valve temperature sensor
3. Supply pressure transducer
4. Flow transducer (Delta pressure transducer and net)
5. Nozzle unit with valve diaphragm
6. Inspiratory solenoid
Air inlet:
Gases from the G
D
AS
ISTRIBUTION
to the inlet tube connectors on the G
tube connectors and the corresponding tubes have
different diametres to prevent faulty gas
connections.
The filter housing and the filter cover is provided with
matching guide pins. These guide pins prevents
mounting the filter cover (with inlet tube connectors)
on wrong G
M
.
AS
ODULE
A non-return valve for the gas inlet is located in the
inlet filter cover. This valve will suppress short
pressure drops in the gas supply. The non-return
valve is also designed to slowly evacuate
compressed gas from the G
supply to the G
M
AS
ODULE
Description of functions
2
3
1
B
are connected
LOCK
M
. The
AS
ODULES
M
if the gas
AS
ODULE
is disconnected.
3 - 7