Siemens KION Service Manual page 59

Anesthesia workstation
Table of Contents

Advertisement

KION / KION-i
E382 E392E 061 01 02 02
Fresh gas pressure container
To remove the fresh gas pressure container:
• Disconnect all tubings from the fresh gas pressure
container.
• Disconnect the position sensor cable.
• Remove the two screws at the base plate.
6
If the fresh gas pressure container is to be
disassembled, note the following:
• Before disassembling the fresh gas pressure
container, mark (1) with a pen the relative positions
between the containers:
– Top cover
– Intermediate parts
– Base plate.
This marking will make a correct assembling easier.
H
• When assembling the fresh gas pressure container,
make sure that the diaphragm (2) and the big O-
rings between the intermediate parts and the base
plate are correctly seated.
The following parts are now accessible inside the
fresh gas pressure container:
• Diaphragm (2)
• PC 1752 Position sensor (3)
• Piston (4)
9
• Piston seal (5). Remove the screw (6) and the
lower part of the piston when replacing the piston
seal.
Note: The piston and PC 1752 are matched and
delivered as a spare part kit. They must be replaced
at the same time.
;
;
<
=
Siemens-Elema AB
Disassembling and assembling
4 - 13