PIVap vaporizers

The PIVap vaporizers must not be disassembled
further than shown below and in the KION Spare
Parts List.
Before disassembling, empty the vaporizer.
• When replacing the window (1), make sure to use
correct window and window frame. Refer to
Service Instructions EM010/02/I.
• To remove the vaporizer valve (2), remove the
screw (3) and carefully pull out the valve.
Note: The screw is sealed with Loctite.
• When mounting the vaporizer valve (2):
– Apply Loctite® 243 on the screw (3).
– Do not tighten the screw (3) to hard, the knob
must be easy to turn.
6
• When replacing the O-ring (4), apply a thin layer of
special grease on the new O-ring.

Vaporizer magazine

To remove the magazine cover (1):
• Remove the three screws (2).
H
• Carefully lift off the magazine cover (1).
To remove the complete vaporizer magazine (3)
6
including magazine cover (the magazine cover can
remain mounted on the magazine):
• Remove the four screws (4).
• Carefully lift off the vaporizer magazine (3).