Siemens KION Service Manual page 162

Anesthesia workstation
Table of Contents

Advertisement

x
409
... 01 99 ...
MEA: DIAG_CAN_BUFFER_FULL
x
410
... 01 9A ...
MEA: DIAG_CAN_DATA_OVERRUN
x
412
... 01 9C ...
MEA: DIAG_CAN_TIMEOUT
x
413
... 01 9D ...
MEA:
DIAG_CAN_WARNING_LIMIT_REACHED
x
415
... 01 9F ...
MEA: DIAG_UNKNOWN_CAN_MESSAGE
x
416
... 01 A0 ...
MEA: DIAG_AD_ERROR_SINGLE_FAULT
Fault: Communication bus (CAN) is not working properly.
Description: The circular CAN receive buffer is full, at least one CAN-message is
missed. Circular buffer holds 10 CAN-messages.
What to do:
1.
If this log is repeted several times; check CAN communicating units, i.e.
microprocessor controlled PC boards, multigas analyzer, KION Monitor and
DuoView. Also check PC board- and cable connectors.
2.
Report to EM-HSC.
Fault: Communication bus (CAN) is not working properly.
Description: Data overrun has occurred, at least one CAN-message is missed.
What to do:
1.
If this log is repeted several times; check CAN communicating units, i.e.
microprocessor controlled PC boards, multigas analyzer, KION Monitor and
DuoView. Also check PC board- and cable connectors.
2.
Report to EM-HSC.
Fault: Communication bus (CAN) is not working properly.
Description: The CAN driver reports a send timeout error.
What to do:
1.
Replace PC 1737 Measuring.
2.
If this log is repeted several times; check CAN communicating units, i.e.
microprocessor controlled PC boards, multigas analyzer, KION Monitor and
DuoView. Also check PC board- and cable connectors.
3.
Report to EM-HSC.
Fault: Communication bus (CAN) is not working properly.
Description: At least one error counter has reached warning limit.
What to do:
1.
If this log is repeted several times; check CAN communicating units, i.e.
microprocessor controlled PC boards, multigas analyzer, KION Monitor and
DuoView. Also check PC board- and cable connectors.
2.
Report to EM-HSC.
Fault: Internal program error e.g. due to incompatible sub-system SW-versions.
Description: Unknown CAN-message is received from an other sub-system.
What to do:
3.
Check sub-system SW-compatibility. Refer to KION Compatibility Chart.
4.
Report to EM-HSC.
Fault: Hardware failure.
Description: Failure to read A/D-port once.
What to do: Replace PC 1737 Measuring.