Siemens KION Service Manual page 128

Anesthesia workstation
Table of Contents

Advertisement

x
076
... 00 4C...
ALA:
DIAG_LOW_BATTERY_CAPACITY_IN_BEGIN
NING_OF_TEST
x
077
... 00 4D...
ALA:
DIAG_BATTERY_VOLTAGE_DECREASING_T
OO_FAST
x
105
... 00 69 ...
CON: DIAG_CAN_BUFFER_FULL
x
106
... 00 6A ...
CON: DIAG_CAN_DATA_OVERRUN
x
109
... 00 6D ...
CON:
DIAG_CAN_WARNING_LIMIT_REACHED
x
112
... 00 70 ...
CON: DIAG_CAN_READ_ERRROR
x
113
... 00 71 ...
CON: DIAG_CAN_UNKNOWN_ MESSAGE
Fault: Low battery capacity.
Description: The internal batteries are below 23.5 V at the beginning of the discharge
test (running on mains or battery).
What to do:
1.
Allow the batteries to charge during 6 hours.
2.
Check / replace internal batteries.
Fault: Battery voltage decreasing too fast.
Description: The internal battery voltage will decrease too fast when running on
battery (more than 0.4 V per minute).
What to do: Replace the internal batteries.
Fault: Communication bus (CAN) is not working.
Description: The circular CAN receive buffer is full, at least one CAN-message is
missed. Circular buffer holds 10 CAN-messages.
What to do:
1.
If this log is repeted several times; check CAN communicating units, i.e.
microprocessor controlled PC boards, multigas analyzer, KION Monitor and
DuoView. Also check PC board- and cable connectors.
2.
Report to EM-HSC.
Fault: Communication bus (CAN) is not working.
Description: Data overrun has occurred, at least one CAN-message is missed.
What to do: Replace PC 1730 Control.
Fault: Communication bus (CAN) is not working.
Description: At least one error counter has reached warning limit.
What to do:
1.
If this log is repeted several times; check CAN communicating units, i.e.
microprocessor controlled PC boards, multigas analyzer, KION Monitor and
DuoView. Also check PC board- and cable connectors.
2.
Report to EM-HSC.
Fault: Internal program error.
Description: SW-related problem. Error reading a CAN-message. Some parameter
was out of range.
What to do: Report to EM-HSC.
Fault: Internal program error e.g. due to incompatible sub-system SW-versions.
Description: Unknown CAN-message is received from an other sub-system.
What to do:
1.
Check sub-system SW-compatibility. Refer to KION Compatibility Chart.
2.
Report to EM-HSC.