Maintenance
General
• Only personnel trained and authorized by Siemens
shall be permitted to perform installation, service or
maintenance of the KION / KION-i workstation.
• A "Twelve months maintenance" must be
performed every twelve months or every 3000
hours of operation whichever comes first. Any
service or maintenance must be noted in the log
book.
• After or in combination with performing the
"Twelve months maintenance" described in this
chapter, a routine cleaning must be performed.
Refer to chapter "Routine cleaning" in the "KION /
KION-i – Operating Manual" for details.
• The internal battery, and the extra internal battery if
fitted, shall be replaced every three years.
• Some equipment used in the KION / KION-i
workstation, but not covered by this Service
Manual, may also demand maintenance actions.
Refer to the equipments documentation. These
units can e. g. be:
– Multigas analyzer
– KION Monitor including its optional equipment
– DuoView
– Suction equipment
– Auxiliary O
supply
2
– Gas evacuation
– Humidifier equipment
– Other optional equipment.
Preparations
7
• Make sure that the KION / KION-i workstation
works properly before performing any maintenance.
• Disconnect the gas supplies (wall and/or tank).
• Disconnect the mains power cable.
• Set the power switch on the control panel to Off.
If the power switch is set in any other position, the
internal battery will supply power to the PC boards.
• Remove patient tubing (incl. gas sampling tube and
water trap) and the manual breathing bag.
• If fitted, remove bacteria filter(s) from the patient
cassette.
7 - 2
Required equipment
• Standard service tools.
• 24 mm socket wrench.
• Pressure Tester 0.2 bar Regulator (REG1).
P/N 65 62 479 E392E.
• Support plate adjustment tool.
P/N 66 37 529 E392E.
• Grease, P/N 95 73 700 E341E. Special grease to be
used in high O
• Loctite® 243.
Handling PC boards
KION workstations delivered before
July 1998 are not equipped with the
protective plate covering the PC
boards inside the pneumatic section.
Avoid contact with these PC boards
when performing maintenance inside
the pneumatic section.
The PC boards contain components that are highly
sensitive to static electricity.
Those who come into contact with circuit boards
containing sensitive components must take certain
precautions to avoid damaging the components (ESD
protection).
When working with ESD sensitive components,
always use a grounded wrist band and grounded
work surface. Adequate service tools must also be
used.
PC boards (spare parts) must always be kept in
protective packaging for sensitive electronic devices.
PC boards must not be inserted or removed while
mains power or battery power is applied to the PC
boards.
Remove and insert the PC boards very carefully to
avoid damage to the connectors.