A


.
E382 E392E 061 01 02 02
?
@

!
"
Siemens-Elema AB
• Open the Power & Communication interface
housing (6).
The Internal battery (7), the AC/DC Converter (8) and
all PC boards inside the Power & Communication
interface are now accessible.
Always disconnect the battery cable (9), when
working inside the Power & Communication
interface.
Keep fingers away when closing the Power &
Communication interface and the Battery drawer!
To remove the internal battery:
• Remove the two screws (10) securing the battery
holder (11).
• Lift off the battery holder and the internal battery.
To remove the AC/DC Converter:
• Remove the internal battery.
• Remove the cable ties and cable clamp securing
the AC/DC Converter cable (12) to the battery
drawer.
• Carefully disconnect the Power cable and Signal
cable connectors (13).
• Remove the seven screws holding the AC/DC
Converter to the battery drawer
• Lift off the AC/DC Converter.
When assembling the unit, make sure that the guide
plate (14) and the rubber seal (15) are mounted on
the AC/DC Converter.