Description of functions
3
MV3
MV4
IN MV2
3 - 20
MV5
OUT
MV1
Siemens-Elema AB
Patient unit
14. Patient cassette
The P
C
is a removable and easy-to-clean
ATIENT
ASSETTE
unit. The five mushroom valves and the three uni-
directional valves in the patient cassette control the
gas flow in the patient unit.
The gas flow is controlled due to:
• Selected breathing system.
• Selected ventilation mode.
• Breath cycle phase.
The expiratory gas flow is measured by the
E
F
T
XPIRATORY
LOW
RANSDUCER
The APL V
is used to limit the airway pressure
ALVE
during manual ventilation.
The P
C
sub-units are:
ATIENT
ASSETTE
• A removable valve assembly plate with five
mushroom valves as indicated in the illustration.
The mushroom valves MV1–5 controls the gas flow
in the P
U
. The mushroom valves are
ATIENT
NIT
controlled by the five solenoid valves EMV1 – 5 on
the S
P
P
UPPORT
LATE FOR
– MV1 controls the gas flow from and to the
M
B
B
ANUAL
REATHING
AG
– MV2 controls the expiratory flow which is
evacuated without being used for rebreathing.
– MV3 controls the gas flow to and from the
B
B
.
REATHING
ELLOWS
– MV4 controls the rebreathing gas flow towards
the patient.
– MV5 controls the fresh gas delivered to the
M
B
B
ANUAL
REATHING
AG
There are also connections for the CO
the valve assembly plate. The absorber
connections (IN and OUT) are equipped with
springloaded cut-off valves that will automatically
close the connections when the CO
lowered.
The valve assembly plate must be replaced during
the "Twelve months maintenance".
KION / KION-i
.
C
.
ATIENT
ASSETTE
.
.
A
BSORBER
2
A
is
BSORBER
2
E382 E392E 061 01 02 02