KION / KION-i
Checking the internal battery
• Allow the internal battery to charge until the battery
indicator on the KION Monitor shows "100%".
• Disconnect mains power to make the KION / KION-
i workstation run on the internal battery.
• Let the KION / KION-i workstation run on the
internal battery and check the battery alarm is not
activated within:
– 40 minutes if the KION / KION-i workstation is
equipped with standard internal battery.
– 130 minutes if the KION / KION-i workstation is
equipped with extra internal battery.
• Allow the internal battery to recharge before clinical
use of the KION / KION-i workstation.
Note: To save time, this battery test can be
performed during the functional checks described
below in section "Completing the Twelve-months
maintenance".
Safety inspection
• Make a visual inspection of the KION / KION-i
workstation for external defects or damages.
• Check the mains power cable, voltage supply
cables and connections for damage.
• Perform a leakage current test. The leakage
current test is a standard procedure regulated by
IEC 60 601-1 or corresponding national standards.
Allowable values and test methods are defined in
the standard. The use of a leakage tester, e. g.
Bender Safety Tester 601/751 or equivalent is
recommended.
• Check that a "KION / KION-i – Operating Manual"
corresponding to the installed software version is
present. Also check that Operating Manuals for all
optional equipment connected to the KION / KION-i
workstation are present.
E382 E392E 061 01 02 02
Completing the Twelve-months maintenance
• Perform a check of all vaporizers to be used on the
KION / KION-i workstation. For details, refer to the
"KION / KION-i – Operating Manual", chapter "Six-
month maintenance", section "Checking the
vaporizers".
• Perform a "Calibration". Refer to the
"KION / KION-i – Operating Manual" for details.
• Perform a "Functional check". Refer to the
"KION / KION-i – Operating Manual" for details.
• Perform "Functional checks" on the optional
equipments connected to the KION / KION-i
workstation. Refer to the Operating Manual for
these equipments (Backup gas supply manifold,
Auxiliary O
supply, Suction equipment, etc).
2
• Note in the log book that a "Twelve-months
maintenance" has been performed.
Siemens-Elema AB
Maintenance
7
7 - 11