Disassembling and assembling
6
H
4
<
4 - 12
9
;
=
Siemens-Elema AB
Inspiratory gas block
Seal for the drive gas module and cuffs for the fresh
gas modules are mounted in the Inspiratory gas
block. To remove the seal/cuffs:
• Remove the gas modules.
• The seal (1) for the drive gas module is now
accessible.
• Remove the three screws holding the cover plate
(2) and lift off the cover plate.
• The cuffs (3) for the fresh gas modules are now
accessible.
Note: Do not use a tool with sharp edges that may
damage the inspiratory gas block when removing the
seal/cuffs.
There are two versions of the one-way valve OV 4:
• Old version (4)
• New version (5).
Note: The O-ring (6) in the valve seat is used only in
combination with OV 4 of the new version (5).