Siemens KION Service Manual page 174

Anesthesia workstation
Table of Contents

Advertisement

Maintenance
7
6
H
9
7 - 8
6
;
Siemens-Elema AB
Dust filter
• Check the dust filter (1) inside the trolley cover.
• Clean the dust filter if it is clogged. The dust filter
can be rinsed in water. Shake out all remaining
water. Make sure that no water remains in the
filter.
Backup gas supply manifold
• On all four yokes, replace:
– Gas inlet filter (1)
– Seal at the gas cylinder connection (2).
• The pressure regulators and the safety valves are
preset to correct settings and secured with a
visible drop of sealant. Inspect the pressure
regulators and the safety valves:
– Check that the seals (3) on the pressure
regulators are not broken.
– Check that the seals (4) on the safety valves are
not broken.
• If a seal on any of these units is broken, replace
this unit.
KION / KION-i
E382 E392E 061 01 02 02