10
Hardware Information
Electrical Connections
Electrical Connections
• The CAN bus is a serial bus with high speed data transfer. The two
• The power input socket accepts a line voltage of 100 – 240 VAC ±10 % with a
Never use cables other than the ones supplied by Agilent Technologies to ensure proper
N O T E
functionality and compliance with safety or EMC regulations.
Serial Number Information (ALL)
The serial number information on the instrument labels provide the following
information:
CCYWWSSSSS
CC
YWW
SSSSS
134
connectors for the CAN bus are used for internal module data transfer and
synchronization.
line frequency of 50 or 60 Hz. Maximum power consumption varies by
module. There is no voltage selector on your module because the power
supply has wide-ranging capability. There are no externally accessible
fuses, because automatic electronic fuses are implemented in the power
supply.
Format
country of manufacturing
• DE = Germany
• JP = Japan
• CN = China
year and week of last major manufacturing change, e.g. 820
could be week 20 of 1998 or 2008
real serial number
Agilent 1290 Infinity Valve Drive and Valve Heads