Introduction to the Column Compartment

Electrical Connections

Electrical Connections
• The CAN bus is a serial bus with high speed data transfer. The two
• One analog output provides signals for integrators or data handling
• The REMOTE connector may be used in combination with other analytical
• With the appropriate software, the RS-232C connector may be used to
• The power input socket accepts a line voltage of 100 – 240 volts AC ± 10%
Never use cables other than the ones supplied by Agilent Technologies to ensure proper
N O T E
functionality and compliance with safety or EMC regulations.
18
connectors for the CAN bus are used for internal Agilent 1200 Series
module data transfer and synchronization.
systems.
instruments from Agilent Technologies if you want to use features such as
start, stop, common shut down, prepare, and so on.
control the module from a computer through a RS-232C connection. This
connector is activated and can be configured with the configuration switch.
See your software documentation for further information.
with a line frequency of 50 or 60 Hz. Maximum power consumption is
220 VA. There is no voltage selector on your module because the power
supply has wide-ranging capability. There are no externally accessible
fuses, because automatic electronic fuses are implemented in the power
supply. The security lever at the power input socket prevents the module
cover from being taken off when line power is still connected.