Agilent Technologies 1290 Infinity User Manual page 58

Valve drive and valve heads
Hide thumbs Also See for 1290 Infinity:
Table of Contents

Advertisement

4
Using the Module
Software Configuration
Assign a Pressure Limit Cluster to the Valve
The valve head's RFID Tag holds the information about its max. pressure. An
assigned pressure cluster partner assures that this limit gets not exceeded.
The max. pressure information of the valve head will become the new
maximum pressure setting of the assigned pressure cluster partner. This helps
to avoid damage to the valve head if the max. pressure of the pump is higher
then the valve head's maximum pressure.
1 Open the valve's control menu (right click on the valve's GUI or via the
2 Assign the valve's pressure partner (also 'None' or several partners
Assigning a Pressure Limit Cluster only works in combination with a 1290 Series Infinity
N O T E
Binary Pump G4220A/B
58
Instrument Tab).
possible).
Agilent 1290 Infinity Valve Drive and Valve Heads