Procedure - Nellcor OxiMax N-65 Operator's Manual

Hendheld pulse oximeter
Hide thumbs Also See for OxiMax N-65:
Table of Contents

Advertisement

When the monitor is turned on, the backlight remains on
and the display cycles through the following sequence as
POST takes place:
All display graphics are shown for three seconds
and the backlight is turned on.
The display goes blank (all display elements off)
for one second.
The software version number shows for 3 seconds
as a three digit number in the right number field
and two dashes in the left number field.
The current time of day is shown in 24 hour
format.
Successful completion of POST is announced by a
POST pass tone. A failed POST is announced by a
high-priority alarm tone.

Procedure

To turn on the monitor:
1. Press the Power button to turn on the monitor.
The backlight remains on throughout POST.
O
Max N-65 Operator's Manual
XI
OXIMAX N-65 Operation
37