Procedure - Nellcor OxiMax NPB-40 Service Manual

Handheld pulse oximeter
Hide thumbs Also See for OxiMax NPB-40:
Table of Contents

Advertisement

Performance Verification

Procedure

16
pigment, and externally applied coloring agents such as nail polish, dye, or
pigmented cream.
Note: The NPB-40 automatically starts the Power-On Self-Test (POST), which
tests the NPB-40 circuitry and functions.
Caution: During POST (immediately after power-up), confirm that all
display segments and icons are shown, and the NPB-40 speaker sounds a
one-second tone.
When the NPB-40 is turned on, the backlight remains on, the display shows the
following sequence, in order, as POST takes place:
Display graphics are shown for three seconds
Display goes blank (all display elements off) for one second
Version number of the software is shown as a three digit number in the
right hand number field (with leading zeros if the version number is less
than 100) and two dashes in the left-hand number field for three seconds
Current time of day, using a 24-hour clock format is shown for three
seconds
Successful completion of the POST is announced by a POST Pass tone.
(A failed POST is announced by a high-priority alarm tone.
1. Turn on the NPB-40 by pressing the Power button.
Note: The backlight remains on during POST.
2. All display numbers and icons are shown for three seconds. The backlight is
turned on.

Advertisement

Table of Contents

Troubleshooting

loading

Table of Contents