Table 6: Electronic Emissions - Nellcor OxiMax NPB-40 Service Manual

Handheld pulse oximeter
Hide thumbs Also See for OxiMax NPB-40:
Table of Contents

Advertisement

NPB-40

Table 6: Electronic Emissions

The NPB-40 is intended for use in the electromagnetic
environment specified below. The customer or user of the NPB-40
should assure that it is used in such an environment.
Emission Test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Electromagnetic
Environment -
Guidance
The NPB-40 uses RF
energy only for its
internal function.
Therefore, the RF
emissions are very low
and not likely to cause
interference in nearby
electronic equipment.
The NPB-40 is suitable
for use in
establishments,
including diagnostic
establishments and
those directly
connected to the public
low-voltage power
supply network that
supplies buildings used
for domestic purposes.
Specifications