Before Applying Power

Verify that the premises electrical supply is within the range of the instrument. The
instrument has an autoranging power supply.
WARNING
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
User's Guide U3042-90002
This Manual:http://www.manuallib.com/agilent/u3042ae08-test-set-user-guide.html
If this product is not used as specified, the protection provided by the
equipment could be impaired. This product must be used in a normal
condition (in which all means for protection are intact) only.
The Mains wiring and connectors shall be compatible with the connector used
in the premise electrical system. Failure, to ensure adequate earth grounding
by not using the correct components may cause product damage, and serious
injury.
Always use the three prong AC power cord supplied with this product. Failure
to ensure adequate earth grounding by not using this cord may cause product
damage and the risk of electrical shock.
This product is designed for use in Installation Category II and Pollution
Degree.
Before switching on this instrument, make sure the supply voltage is in the
specified range.
Verify that the premise electrical voltage supply is within the range specified
on the instrument.
Ventilation Requirements: When installing the instrument in a cabinet, the
convection into and out of the instrument must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the instrument by 4 °C for every 100 watts dissipated in the
cabinet. If the total power dissipated in the cabinet is greater than 800 watts,
forced convection must be used.
Safety and Information
63