N5242A Rf Interface Cable Connections - Agilent Technologies U3042AE08 User Manual

Table of Contents

Advertisement

U3042AE08

N5242A RF Interface Cable Connections

Figure 17 on page 20
configured to the PNA-X. The cables have been supplied with U3021PL3 Option 442.
1. Connect the RF interconnect cables from the PNA-X to the test set in the order listed.
As you are connecting each cable, torque to 8 in-lb. The longer, straight end of each
cable is connected to the test set. Refer to
CAUTION
Table 8
Numeric
Order
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
User's Guide U3042-90002
This Manual:http://www.manuallib.com/agilent/u3042ae08-test-set-user-guide.html
illustrates the setup configuration of the test set and how it should be
Over torque will cause damage to the test set and may cause connectors to
spin or become loose.
N5242A RF Interface Cable Connection
RF Cables
From: PNA-X
U3042-20031
Port 1 SOURCE OUT
U3042-20032
Port 1 CPLR THRU
U3042-20029
Port 1 CPLR ARM
U3042-20030
Port 1 RCVR A IN
U3042-20035
Port 3 SOURCE OUT
U3042-20033
Port 3 CPLR ARM
U3042-20034
Port 3 RCVR C IN
U3042-20036
Port 3 CPLR THRU
U3042-20039
Port 4 SOURCE OUT
U3042-20037
Port 4 CPLR ARM
U3042-20038
Port 4 RCVR D IN
U3042-20040
Port 4 CPLR THRU
U3042-20041
Port 2 CPLR ARM
U3042-20044
Port 2 CPLR THRU
U3042-20043
Port 2 SOURCE OUT
U3042-20042
Port 2 RCVR B IN
System Setup with N5242A
Table 8
and
Figure 17 on page
To: Test Set
Port 5 SOURCE IN
Port 5 CPLR THRU
Port 5 CPLR ARM
Port 5 RCVR OUT
Port 6 SOURCE IN
Port 6 CPLR ARM
Port 6 RCVR OUT
Port 6 CPLR THRU
Port 7 SOURCE IN
Port 7 CPLR ARM
Port 7 RCVR OUT
Port 7 CPLR THRU
Port 8 CPLR ARM
Port 8 CPLR THRU
Port 8 SOURCE IN
Port 8 RCVR OUT
20.
19