Environmental Requirements

Refer to your analyzer's standard documentation for environmental requirements. The
PNA-L provides front panel access to the source outputs, receiver inputs and couplers for
use with multiport test sets.

Environmental Tests

The test set complies with all applicable safety and regulatory requirements for the
intended location of use.
• Pressure Altitude (Operation)
3,000 meters (~10,000 feet)
• The instrument can safely operate in a relative humidity of 80% for temperatures to
31 degrees C, decreasing linearly to 50% relative humidity at 40 degrees C.

Equipment Heating and Cooling

If necessary, install air conditioning and heating to maintain the ambient temperature
within the appropriate range. Air conditioning capacity must be consistent with the BTU
ratings given in
CAUTION

Required Conditions for Accuracy Enhanced Measurement

Accuracy–enhanced (error–corrected) measurements require the ambient temperature of
the analyzer and test set to be maintained within ± 1 °C of the ambient temperature at
calibration.

Dimensions and Space Requirements

Standard installation of the test set and analyzer includes configuration and installation on
a customer provided lab bench or table top of adequate size and strength. For weight,
dimensions and space requirements, refer to the network analyzer documentation that is
used to configure the test set.
Table 2
Model
U3042AE08
6
This Manual:http://www.manuallib.com/agilent/u3042ae08-test-set-user-guide.html
Table
1.
Ventilation Requirements: When installing the instrument in a cabinet, the
convection into and out of the instrument must not be restricted. The ambient
temperature (outside the cabinet) must be less than the maximum operating
temperature of the instrument by 4 °C for every 100 watts dissipated in the
cabinet. If the total power dissipated in the cabinet is greater than 800 watts,
forced convection must be used.
Instrument Dimensions
Weight
11.4 kg (25 lb)
Height
Width
Depth
19.1 cm
42.5 cm
43.2 cm
(7.5 in)
(16.7 in)
(17 in)
General Specifications
User's Guide U3042-90002