Figure 48
10.Compare the Reflection Tracking (1,1) trace to the appropriate limits in
page 45
(press Marker/Analysis > Analysis > Limit Test). The trace should be above the limit.
PASS will be displayed on the screen if the limit lines are used.
11.Repeat
Figure 49
NOTE
52
This Manual:http://www.manuallib.com/agilent/u3042ae08-test-set-user-guide.html
Option 001 or 002 Reflection Tracking Trace (Port 1-4) with N5242A
(N5230C) or
Table 15 on page 46
step 9
and
step 10
for Cal Sets "999.1" thru "999.12" (12-Port).
Option 001 or 002 Reflection Tracking Trace (Port 5-12) with
N5242A
Response from 10 MHz to 500 MHz is normal due to the PNA-X Couplers in
comparison to the test set bridges. The bridges have more gain in the coupled
(N5242A). This can be done using Limit Lines
User's Guide U3042-90002