N5230C or N5242A 1-Port Calibration and Verification Procedure

1. Connect the ECal or the mechanical cal kit to Port 1 or the port you are testing. Refer to
Figure 2 on page
Figure
2. Perform a 1-Port Calibration on Port 1. On the analyzer, press Response > CAL > Start
Cal > Calibration Wizard.
a. If using a mechanical cal kit, select SmartCal.
b. If using an ECal module, select ECal.
3. Continue following the Cal Wizard prompts. In the "Select Calibration Ports and ECal
Module" window, press Clear All and select Port 1, then press Next.
Figure 42
4. Continue to follow the prompts until the "Calibration Completed" window appears.
5. At the Calibration Completed prompt, select Save As User Calset and type the name
999.1. Overwrite the Calset if it already exists. Press Save.
NOTE
48
This Manual:http://www.manuallib.com/agilent/u3042ae08-test-set-user-guide.html
2. Torque to 8 in-lb. For more information press the Help button, see
42.
1-Port Calibration
If you do not have a key board, select Save As User Calset > Edit Name and
save as 999.x. X is the port number you are calibrating. See
the numeric keypad on the analyzer's front panel to enter "999.1."
Operational Check
Figure
43. Use
User's Guide U3042-90002