Agilent Technologies U3042AE08 User Manual page 43

Table of Contents

Advertisement

U3042AE08
Controlling the Test Set with N5230C or N5242A
Channel 1:
Specifies the channel number for dialog settings. Each channel is configured individually.
The list box illustrates the channels that currently have measurements. There must be at
least one measurement present in order to make the settings.
Channel Control Label:
Specifies the label to be displayed on the screen during the channel sweep.
Before Sweep Start– After Sweep End Tabs:
Commands /data for all four interfaces can be sent Before Sweep Start and After Sweep
End. However, they are configured and enabled on separate tabs of the Interface Control
dialog box. For example; to send GPIB commands Before and After a sweep, the Enable
Control check box must be selected and commands entered on both the Before Sweep Start
and After Sweep End tabs.
Before Sweep Start:
The data is sent before the first trace on the channel begins sweeping.
After Sweep End:
The data is sent after the last trace on the channel completes sweeping.
Address:
Positive integer is used to specify switch path to be controlled. Address also specifies
output on specific ports. Refer to
"Test Set I/O Interface Commands" on page
43.
Dwell After Command:
Specifies a wait time, in milliseconds, after all commands to all interfaces are sent. Any
positive integer is allowed. This is used to allow all external devices to settle before
beginning a measurement. An erratic trace could indicate that more settling time is
necessary.
Reset All:
Sets all fields on all channels to their default values.
Handler I/O Control and Aux I/O Output Voltage:
Provides I/O interface control through the rear panel of the analyzer. Refer to the Help
menu for further information.
Save and Recall:
Saves and recalls the contents of the dialog box. If the Interface Control dialog box is
populated with settings during an Instrument State Save, the settings are automatically
recalled with the instrument state settings. Interface control uses an *.xml file type. An
example file is stored on the analyzer's hard drive. You can recall it into the dialog, or you
can open and edit it with a word processor, such as Word Pad.
34
User's Guide U3042-90002
This Manual:http://www.manuallib.com/agilent/u3042ae08-test-set-user-guide.html