X-Ray Group (40 Kw Configuration) - AGFA DR 100s User Manual

Table of Contents

Advertisement

X-ray group (40 kW configuration)

Table 31: X-ray specifications
Nominal power (IEC 60601-1)
kV values
Inverter frequency in high voltage
mA values
mAs values
Exposure times
Table 32: Accuracy (IEC 60601-2-54)
kV
mA
mAs
Time
Table 33: Equivalent filtration @75 kV
Filtration
Collimator
Total filtration
Additional filtration of DAP meter (optional)
Total filtration with DAP meter (optional)
±5%
±20%
±(10% + 0.2 mAs)
±(10% + 0.1 ms)
DR 100s | Product Information | 179
40 kW @ 100 kV, 400 mA
40 - 150 kV at steps of 1 kV
40 kHz
70 - 400 mA
(automatically associated to kV)
0.1 - 500 mAs
(with increases of 12.5%)
0.001 - 3 s (DR mode)
0.001 - 10 s (CR mode)
(according to mAs)
0411D EN 20220627 1040
1.5 mm Al
2 mm Al
3.5 mm Al
0.3 mm Al
3.8 mm Al

Advertisement