152 | DR 100s | Problem solving

The equipment cannot be switched off

Details
Cause
Brief
Solu-
tion
0411D EN 20220627 1040
The procedure to stop DR 100s does not succeed. The power status
indicator light remains on.
Wait until the complete shutdown procedure has finished. This
may take up to 5 minutes.
Failure of shutdown procedure.
Insert a pointed object (for example a ball pen) in the hole placed
under the console. Press and hold for ten seconds in order to force
the equipment shutdown.
Figure 67: Button to force a shutdown
If the equipment still does not switch off, place it in a suitable
place, proceed with charging the battery and contact your local
service organization for assistance.