Automatic collimator

Model
Manufacturer
Collimation
Coverage of the field at
100cm SID
Light field accuracy
Lighting source
Light intensity (IEC
60601-2-54)
Minimum contrast ratio (IEC
60601-2-54)
Focal distance measurement
Maximum radiation leakage
Inherent filtration
Additional filtration
Rotation
Weight
DR 100s | Product Information | 177
R221 ACS
Ralco
Via dei Tigli 13/G
20853 Biassono (MB), Italy
Motorized (and manual) with internal light
source, multilayer, squared field, double la-
ser at overlaid lines for SID control at 100 cm
adjustable by service up to 190 cm.
minimum 0 cm x 0 cm
maximum 48 cm x 48 cm
< 2% of the used SID
Clusters of high-brightness LED power
> 160 lux
4:1
Retractable tape measure (maximum exten-
sion 3 m)
150 kVp - 4mA
2 mm equivalent Al at 75kV
Motorized selection
• 0 mm Al
• 1 mm Al + 0.1 mm Cu
• 1 mm Al + 0.2 mm Cu
• 2 mm Al
± 90°
(integrated handles with pushbutton to re-
lease rotation brake)
8 kg
0411D EN 20220627 1040