Step 1 - Program - Teledyne 4250 Installation And Operation Manual

Area velocity flow meter
Table of Contents

Advertisement

4250 Flow Meter
Section 2 Programming

2.4.3 Step 1 - Program

2-28
If you continue to work through the rest of the program, the lock
will not engage until you are done. But if you stop programming
longer than two minutes, the lock will engage, and you will not be
able to make any further program changes.
Press Enter and the SETUP menu will reappear.
SELECT OPTION
• PROGRAM • • SETUP •
PROGRAM will flash. (PROGRAM is always the default choice.
That is because you are more likely to need to make changes in
the PROGRAM section of the software than in the SETUP
section.)
Note
If you choose NOT MEASURED for any selection, the flow
meter will make no further reference to that function for the rest
of the program, and you will be unable to activate that function
later. If there is a feature you need that does not appear when
the manual says it should, return to step 1 and make sure you
have not accidentally left it turned off.
If you program a parameter value as a condition for sampler
enabling, pacing, dialout, etc., and then turn that parameter
sensor off, the flow meter will also remove that condition from
the program.
Consider all aspects of your program before you make any
changes.
Press Enter. The following will appear:
UNITS OF LEVEL MEASUREMENT
• FT • • IN • • M • • MM • • NOT MEASURED •
Selection of feet, inches, meters, or mm depends on your situ-
ation. You would select NOT MEASURED if you were using the
flow meter for some other form of sensing only, such as pH, or
temperature. Press Enter. The following will appear. You will
have to press the right arrow key several times to see all of the
options displayed in the following menu:
FLOW RATE UNITS OF MEASURE
• GPS • • GPM • • GPH • • MGD • • CFS • • CFM • • CF-
Pressing the right arrow key several times will move other
units onto the display.:
FLOW RATE UNITS OF MEASURE
• •CFH • • CFD • • LPS • • M3S • • M3M • • M3H • • M3D • • AFD •
NOT MEASURED will also appear if you keep moving with the
right arrow key. This step establishes the units of measure the
flow meter will use in all subsequent displays and calculations.