Calibrating The D.o. Probe With A Flow Meter - Teledyne 4250 Installation And Operation Manual

Area velocity flow meter
Table of Contents

Advertisement

4250 Flow Meter
Section 4 Options and Accessories
Figure 4-10 D.O. Parameter Module
4.11.6 Calibrating the D.O.
Probe with a Flow
Meter
4-26
You must use the Isco 270 D.O. Module box between the probe
and flow meter; this extends the distance between the probe and
the flow meter to 1,000 feet.
Prepare the probe as described above if this has not already been
done (fill the probe cavity with electrolyte and seat the mem-
brane).
Note
You must use the Isco Temperature Probe with the D.O. Probe
to provide temperature compensation.
Wrap both the D.O. Probe and Temperature Probe in a damp
cloth. Wait ten minutes for it to stabilize, then proceed.
Go to step 1 on the flow meter. Select PROGRAM, then step
through the units of measure with Enter until you reach the
menu for pH measurement. Select NOT MEASURED. Press
Enter. Then D.O. UNITS will appear. Select either MG/L or
PPM. Press Enter. Exit the program and re-enter going to step
3.
You must select measurement of D.O. in step 1 or the D.O. menu
will not appear in step 3. If you cannot get the D.O. menu to
appear in step 3, recheck your selections in step 1. Remember
that you must select NOT MEASURED for pH in step 1 or D.O.
will not appear on the menu for the rest of the program.
SELECT PARAMETER TO ADJUST
NONE LEVEL D.O.
Select D.O. Press Enter. This display will appear:
D.O. CALIBRATION
D.O STANDARD, ABS BAROMETRIC PRESSURE,