2.4.10 Step 7 - Alarm Dialout
Mode
Select YES to reset the sampler enable feature; select NO to
leave the sampler enabled.
PRINTER ON/OFF WITH ENABLE
• YES • • NO •
This selection allows you to turn the flow meter's internal printer
on or off when the sampler is enabled from the flow meter. This
allows you to conserve battery power and only print a chart when
the sampler is enabled. This feature is useful for monitoring
storm water runoff.
This step lets you signal an alarm to a remote location from the
flow meter. The conditions that can cause an alarm are the same
as described previously for sampler enabling.
Note
You must have the optional internal modem installed and con-
nected to a telephone network to make use of this feature.
This menu will not even appear if you do not have a
modem. The flow meter will automatically advance to the next
step.
More information about the modem can be found in Section 4. If
you have the modem installed and select step 7, the following
will appear:
ALARM DIAL OUT
• DISABLE • • CONDITIONAL • • STORM • • FLOWLINK •
If you select DISABLE, this option will be inactivated until you
change the selection later. The program will advance to the next
step. If you select STORM, the flow meter will request definitions
similar to those for STORM in sampler enable. The following will
appear:
LEVEL
GREATER THAN X.XXX FT (or other units of measure)
Then:
RAINFALL AMOUNT
X.XX INCHES (or other units of measure)
Then:
RAINFALL TIME PERIOD
• 15 MIN • • 30 MIN • • 1 HR • • 2 HR • • 4 HR •
If you press the right arrow key several times, the following
times will appear:
RAINFALL TIME PERIOD
• 6 HR • • 8 HR • • 12 HR • • 24 HR • • 48 HR • • 72 HR •
4250 Flow Meter
Section 2 Programming
2-55